Specifications
Typical for -40 to 70 °C unless otherwise noted.
Input Characteristics
Number of inputs ................................ 8 differential
Resolution ........................................... 16 bits, 1 in 65,536
Type of ADC ........................................ Delta-sigma
Filters................................................... 15, 60, 240 Hz, user-selectable
Excitation voltage ............................... 2.5, 5, 10 V, user-selectable,
current-limited to 21 mA
per channel
Bridge configurations.......................... Half- and full-bridge circuits,
quarter-bridge with external
completion resistor
Gage factor ......................................... Any (software value)
Input impedance ................................. 20 M
Offset error
15 to 35 ˚C ...................................... 7.6 µV typ, 28 µV max
-40 to 70 ˚C..................................... 50 µV typ, 140 µV max
Gain error
15 to 35 ˚C ...................................... 0.06%
-40 to 70 ˚C..................................... 0.4%
Input noise (60 Hz filter)
±3.5 mV/V ....................................... ±3 LSB
pp
±7.5 mV/V ....................................... ±2 LSB
pp
Other ranges ....................................... ±1 LSB
pp
All-channel update periods
15 Hz ............................................... 1.15 s
60 Hz ............................................... 0.95 s
240 Hz ............................................. 0.90 s
Isolation Voltage
Maximum isolation voltage ................ 250 V
rms
, Installation Category II
Channel-to-channel isolation.............. No isolation between channels
Transient overvoltage.......................... 2,300 V
rms
Physical Characteristics
LED indicators
POWER (green) ............................... Power on and self-test passed
READY (gr
een)
................................
Module configur
ed and r
eady
OVERCURRENT (r
ed)....................... Overcurrent condition on channel
Dimensions
(including terminal base) .................... 10.9 by 10.7 by 9.1 cm
(4.3 by 4.2 by 3.6 in.)
Weight................................................. 110 g (3.7 oz)
Power Requirement
Power from network module .............. 1,000 mW
Envir
onmental
Operating temperatur
e ....................... -40 to 60 °C
Storage temperature........................... -55 to 85 °C
Relative humidity ................................
10 to 90%, noncondensing
Maximum altitude............................... 2,000 m; at higher altitudes the
isolation voltage ratings must
be lower
ed
Pollution degree.................................. 2
Shock and Vibration
N
I recommends Compact FieldPoint if your
application is subject to shock and vibration.
Operating vibration, random
(IEC 60068-2-64) ............................. 10 to 500 Hz, 5 g
rms
Operating vibration, sinusoidal
(IEC 60068-2-6) ............................... 10 to 500 Hz, 5 g
Operating shock
(IEC 60068-2-27) ............................. 50 g, 3 ms half sine,
18 shocks at 6 orientations;
30 g, 11 ms half sine,
18 shocks at 6 orientations
Safety and Compliance
Safety
This pr
oduct is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
IEC 61010-1, EN 61010-1
UL 61010-1, CSA 61010-1
Note: For UL and other safety certifications, refer to the product label or
visit
ni.com/certification, search by model number or product line, and
click the appropriate link in the Certification column.
Electr
omagnetic Compatibility
This product is designed to meet the requirements of the following
standards of EMC for electrical equipment for measurement, control,
and laboratory use:
EN 61326 EMC requirements; Minimum Immunity
EN 55011 Emissions; Group 1, Class A
CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A
Note: For EMC compliance, operate this device accor
ding to
product documentation.
CE Compliance
This product meets the essential requirements of applicable European
Dir
ectives, as amended for CE marking, as follows:
2006/95/EC; Low-Voltage Directive (safety)
2004/108/EC; Electr
omagnetic Compatibility Dir
ective (EMC)
Note: Refer to the Declaration of Conformity (DoC) for this product for
any additional r
egulator
y compliance infor
mation. T
o obtain the DoC for
this pr
oduct, visit
ni.com/cer
tification
, sear
ch by model number or
product line, and click the appropriate link in the Certification column.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers: At the end of their life cycle, all products must be
sent to a WEEE recycling center. For more information about WEEE
recycling centers and National Instruments WEEE initiatives,
visit
ni.com/environment/weee.htm.
Strain Gage Module for Compact FieldPoint
3
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)
Seitenansicht 2
1 2 3 4

Kommentare zu diesen Handbüchern

Keine Kommentare